|
The Global Leader of Portable Data Acquisition Solutions
|
|
1-888-714-3272 |
IOtech Events |
| Selection Guides | OEM Zone |
|
Chapter 6 THERMOCOUPLE BASICS Equation 6.01. Seebeck Voltage
Thermocouple junctions alone do not generate voltages. The voltage or potential difference that develops at the output (open) end is a function of both the temperature of the junction T1 and the temperature of the open end T1. T1 must be held at a constant temperature, such as 0°C, to ensure that the open end voltage changes in proportion to the temperature changes in T1. In principle, a TC can be made from any two dissimilar metals such as nickel and iron. In practice, however, only a few TC types have become standard because their temperature coefficients are highly repeatable, they are rugged, and they output relatively large voltages. The most common thermocouple types are called J, K, T, and E, followed by N28, N14, S, R, and B (See the table in Figure 6.02). In theory, the junction temperature can be inferred from the Seebeck voltage by consulting standard tables. In practice, however, this voltage cannot be used directly because the thermocouple wire connection to the copper terminal at the measurement device itself constitutes a thermocouple junction (unless the TC lead is also copper) and outputs another emf that must be compensated. Cold-Junction Compensation The above example is a special case, however, because one lead of the type-T thermocouple is copper. A constantan/iron thermocouple, on the other hand, needs further consideration (See Figure 6.04). Here, J2 in the ice bath is held constant, and J1 measures the environment. Although J3 and J4 are effectively thermocouple junctions, they are at the same temperature on the isothermal block, so they output equal and opposite voltages and thus cancel. The net voltage is then the thermocouple J1 output representing T1, calibrated to the NIST standard table. If the I/O block were not isothermal, copper wire leads would be added between the input terminal and the copper/iron leads, and the copper/iron junctions (J3 and J4) would be held in an ice bath as well, as illustrated in Figure 6.05. Software Compensation When the ice baths are eliminated, cold junction compensation (CJC) is still necessary, however, in order to obtain accurate thermocouple measurements. The software has to read the isothermal block temperature. One technique widely used is a thermistor, mounted close to the isothermal terminal block that connects to the external thermocouple leads. No temperature gradients are allowed in the region containing the thermistor and terminals (See Figure 6.06). The type of thermocouple employed is preprogrammed for its respective channel, and the dynamic input data for the software includes the isothermal block temperature and the measured environmental temperature. The software uses the isothermal block temperature and type of thermocouple to look up the value of the measured temperature corresponding to its voltage in a table, or it calculates the temperature more quickly with a polynomial equation. The method allows numerous channels of thermocouples of various types to be connected simultaneously while the computer handles all the conversions automatically. Hardware Compensation Type Mixing Linearization The standard method for obtaining high conversion accuracy for any temperature uses the value of the measured thermocouple voltage plugged into a characteristic equation for that particular type thermocouple. The equation is a polynomial with an order of six to ten. The NIST table in Figure 5.32 lists the polynomial coefficients for some common thermocouples. The computer automatically handles the calculation, but high-order polynomials take significant time to process. In order to accelerate the calculation, the thermocouple characteristic curve is divided into several segments. Each segment is then approximated by a lower order polynomial. Analog circuits are employed occasionally to linearize the curves, but when the polynomial method is not used, the thermocouple output frequently connects to the input of an ADC where the correct voltage to temperature match is obtained from a table stored in the computers memory. For example, one data acquisition system TC card includes a software driver that contains a temperature conversion library. It changes raw binary TC channels and CJC information into temperature readings. Some software packages for data acquisition systems supply CJC information and automatically linearize the thermocouples connected to the system. THERMOCOUPLE MEASUREMENT PITFALLS Additional Concerns Thermocouples can become un-calibrated and indicate the wrong temperature when the physical makeup of the wire is altered. Then it cannot meet the NIST standards. The change can come from a variety of sources, including exposure to temperature extremes, cold working the metal, stress placed on the cable when installed, vibration, or temperature gradients. The output of the thermocouple also can change when its insulation resistance decreases as the temperature increases. The change is exponential and can produce a leakage resistance so low that it bypasses an open-thermocouple wire detector circuit. In high-temperature applications using thin thermocouple wire, the insulation can degrade to the point of forming a virtual junction as illustrated in Figure 6.09. The data acquisition system will then measure the output voltage of the virtual junction at T1 instead of the true junction at T2. In addition, high temperatures can release impurities and chemicals within the thermocouple wire insulation that diffuse into the thermocouple metal and change its characteristics. Then, the temperature vs. voltage relationship deviates from the published values. Choose protective insulation intended for high-temperature operation to minimize these problems. Thermocouple Isolation Auto-Zero Correction One effective method of subtracting the offset due to drift is done in two steps. First, the internal channel sequencer switches to a reference node and stores the offset error voltage on a capacitor. Next, as the thermocouple channel switches onto the analog path, the stored error voltage is applied to the offset correction input of a differential amplifier and automatically nulls out the offset (See Figure 6.10). Open Thermocouple Detection Galvanic Action Thermal Shunting Despite these advantages, extension wire generally operates over a much narrower temperature range and is more likely to receive mechanical stress. For these reasons, the temperature gradient across the extension wire should be kept to a minimum to ensure accurate temperature measurements. Improving Wire Calibration Accuracy RTD MEASUREMENTS Platinum RTD resistances range from about 10 W for a birdcage configuration to 10 kW for a film type, but the most common is 100 W at 0°C. Commercial platinum wire has a standard temperature coefficient, a, of 0.00385 W/W/°C, and chemically pure platinum has a coefficient of 0.00392 W/W/°C. The following equation shows the relationship between the sensors relative change in resistance with a change in temperature at a specific a and nominal sensor resistance. Equation 6.02. RTD Temperature Coefficient
A nominal 100-W platinum wire at 0°C will change resistance, either plus or minus, over a slope of 0.385 W/°C. For example, a 10°C rise in temperature will change the output of the sensor from 100 W to 103.85 W, and a 10°C fall in temperature will change the RTD resistance to 96.15 W. Because RTD sensor resistances and temperature coefficients are relatively small, lead wires with a resistance as low as ten ohms and relatively high temperature coefficients can change the data acquisition systems calibration. The lead wires resistance change over temperature can add to or subtract from the RTD sensors output and produce appreciable errors in temperature measurement. The resistance of the RTD (or any resistor) is determined by passing a measured current through it from a known voltage source. The resistance is then calculated using Ohms Law. To eliminate the measurement error contributed by lead wires, a second set of voltage sensing leads should be connected to the sensors terminals and the opposite ends connected to corresponding sense terminals at the signal conditioner. This is called a four-wire RTD measurement. The sensor voltage is measured directly and eliminates the voltage drop in the current carrying leads. Measurement Approaches Figure 6.12 shows a basic two-wire resistance measurement method. The RTD resistance is measured directly from the Ohmmeter. But this connection is rarely used since the lead wire resistance and temperature coefficient must be known. Often, both properties are not known, nor are they convenient to measure when setting up a test. Figure 6.13 shows a basic four-wire measurement method using a current source. The RTD resistance is V/A. This connection is more accurate than the two-wire method, but it requires a high stability current source and four lead wires. Because the high-impedance voltmeter does not draw appreciable current, the voltage across the RDT equals Vm. Equation 6.03. 4-Wire RTD With Current Source
Figure 6.14 shows a three-wire measurement technique using a current source. The symbols Va and Vb represent two voltages measured by the high-impedance voltmeter in sequence through switches (or a MUX), S1 and S2. The RTD resistance is derived from Kirchhoffs voltage law and by solving two simultaneous equations. (Illustrating the solution is beyond the scope of this book.) The benefit of this connection over that shown in Figure 6.13 is one less lead wire. However, this connection assumes that the two current-carrying wires have the same resistance. Equation 6.04. 3-Wire RTD With Current Source
Figure 6.15 shows a four-wire measurement system using a voltage source. The RTD resistance also is derived from Kirchhoffs voltage law and four simultaneous equations based on the four voltages, Va through Vd. The voltage source in this circuit can vary somewhat as long as the sense resistor remains stable. Equation 6.05. 4-Wire RTD With Voltage Source
Figure 6.16 shows a three-wire measurement technique using a voltage source. The RTD resistance is derived from Kirchhoffs voltage law and three simultaneous equations. The voltage source can vary as long as the sense resistor remains stable, and the circuit is accurate as long as the resistances of the two current-carrying wires are the same. Equation 6.06. 3-Wire RTD With Voltage Source
The RTD output is more linear than the thermocouple, but its range is smaller. The Callendar-Van Dusen equation is often used to calculate the RTD resistance: Equation 6.07. RTD Curve Fitting
An alternative method involves measuring RTD resistances at four temperatures and solving a 20th order polynomial equation with these values. It provides more precise data than does the a, d, and b coefficients in the Callendar-Van Dusen equation. The plot of the polynomial equation in Figure 6.17 shows the RTD to be more linear than the thermocouple when used below 800°C, the maximum temperature for RTDs. Self-Heating A typical value for self-heating error is 1°C/mW in free air. An RTD immersed in a thermally conductive medium distributes this heat to the medium and the resulting error is smaller. The same RTD rises 0.1°C/mW in air flowing at one m/s. Using the minimum excitation current that provides the desired resolution, and using the largest physically practical RTD will help reduce self-heating errors. Scanning Inputs Two practical methods for scanning an RTD include constant current and ratiometric. An example of a constant current circuit is shown in Figure 6.18. Its an RTD scanning module, which switches a single 500 µA constant current source among 16 channels. A series of front-end multiplexers direct the current to each channel sequentially while the measurement is being taken. Both three and four wire connections are supported to accommodate both types of RTDs. By applying current to one RTD at a time, errors due to resistive heating become negligible. Advantages of the constant current method include simple circuits and noise immunity. But the disadvantage is the high cost of buying or building an extremely stable constant current source. By contrast, the ratiometric method uses a constant voltage source to provide a current, Is, through the RTD and a resistor, Rd. Four voltage readings are taken for each RTD channel, Va, Vb, Vc, and Vd (See Figure 6.19). The current, voltage, and resistance of the RTD is: Equation 6.08. 4-Wire RTD Ratiometric Measurement
For a three-wire connection (Figure 6.20), the voltage, Va Vc, includes the voltage drop across only one lead. Because the two extension wires to the transducer are made of the same metal, assume that the drop in the first wire is equal to the drop in the second wire. Therefore, the voltage across the RTD and its resistance is: Equation 6.09. 3-Wire RTD Ratiometric Measurement
Practical Precautions THERMISTOR MEASUREMENTS Thermistors are generally composed of semiconductor materials or oxides of common elements such as cobalt, copper, iron, manganese, magnesium, nickel, and others. They typically come with 3 to 6-in. leads, encapsulated, and color-coded. They are available in a range of accuracies from ±15°C to ±1°C, with a nominal resistance ranging from 2,000 to 10,000 W at 25°C. A value of 2252 W is common and can be used with most instruments. A plot of the temperature vs. resistance characteristic curves is usually provided with the device to determine the temperature from a known resistance. However, the devices are highly non-linear and the following equation may be used to calculate the temperature: Equation 6.10. Thermistor Temperature
The constants A, B, and C are calculated from three simultaneous equations with known data sets: Insert R1 and T1; R2 and T2; R3 and T3, then solve for A, B, and C. Interpolation yields a solution accurate to ±0.01°C or better. Linearization Equation 6.11. Thermistor Voltage Divider
If R is a thermistor, and the output voltage is plotted against the temperature, the curve resembles an S-shape with a fairly straight center portion. However, adding other resistors or thermistors to R linearizes the center portion of the curve over a wider temperature range. The linear section follows the equation of a straight line, Y = mX +b: Equation 6.12. Thermistor Voltage Mode
For the resistance mode, see Figure 6.23. Equation 6.13. Thermistor Resistance Mode
Although a lot of research has gone into developing linear thermistors, most modern data acquisition system controllers and software handle the linearization, which makes hardware linearization methods virtually obsolete. Stability Time Constant Dissipation Factor Tolerance Curves Measurement Approaches Differential thermometers: Two thermistors can be used in a Wheatstone bridge to accurately measure the difference in temperature between them. Thermistors can be attached to any heat conducting medium in a system at various points to measure the temperature gradient along its length. Two or more thermistors may be placed in a room to measure temperatures at several different elevations using the same basic switching arrangement. ...to read the entire 144-page book, order your copy today!
|
![]() |
|
[ IOtech PC-Based Data Acquisition (DAQ) and Instrumentation - PDF Format IOtech PC-based products (USB, PCI, Ethernet, GPIB) and signal conditioning options capture waveforms and measure most physical parameters including temperature, vibration, strain, velocity, acceleration, position, as well as common voltage, current, power, and data logging. IOtech's solutions are used in environments such as |