|
The Global Leader of Portable Data Acquisition Solutions
|
|
1-888-714-3272 |
IOtech Events |
| Selection Guides | OEM Zone |
|
Chapter 7 STRAIN BASICS STRAIN MEASUREMENT CONFIGURATIONS Full-Bridge Circuits A full bridge contains four strain gages mounted on a test member (See Figure 7.01). Two gages are mounted on the surface under tension and the other two are mounted on the opposite surface under compression. As the member deflects, the two gages in tension increase in resistance while the other two decrease, unbalancing the bridge and producing an output proportional to the displacement. The bridge output voltage is given by: EQN: 7.01. Full-Bridge Output Voltage
The bridge nulls out potential error factors such as temperature changes because all four strain gages have the same temperature coefficient and are located in close proximity on the specimen. The resistance of the lead wire does not affect the accuracy of the measurement as long as the input amplifier has high input impedance. For example, an amplifier with a 100-MW input impedance produces negligible current flow through the measurement leads, minimizing voltage drops due to lead resistance. Half-Bridge Circuits EQN. 7.02. Half-Bridge Output Voltage
For a large DR, half bridge and quarter-bridge circuits can introduce an additional nonlinearity error (See Figure 7.02). Also, the readings are not accurate when the temperature coefficients between bridge completion resistors and strain gages are different and the resistances do not change proportionally with temperature. Furthermore, bridge completion resistors are not usually located near the strain gages, so temperature differences contribute additional errors. In systems with long lead wires, the bridge completion resistors should be attached close to the gages; however, this may not always be practical due to test fixture limitations or other physical conditions. Quarter-Bridge Circuits EQN. 7.03. Quarter-Bridge Output Voltage
This arrangement has the smallest output, so noise is a potential problem. Furthermore, all the error sources and limitations in the half-bridge apply to the quarter bridge circuit (See Figure 7.03). Excitation Source An ideal data acquisition system provides an excitation source for each channel, independently adjustable from 0.5 to 10.5 V with a current limit of 100 mA. An excitation voltage, V, used with a strain gage of resistance, R, requires a current of I = V/R. The resistance of a Wheatstone bridge measured between any two symmetrical terminals equals the value of one of the resistance arms. For example, four 350-W arms make a 350-W bridge. The load current equals the excitation voltage divided by the bridge resistance; in this case, 10 V/350 = 0.029 A = 29 mA. Heating Consider a Kelvin connection for applying the excitation voltage. Because the excitation leads carry a small current, they drop a correspondingly small voltage; V = I/RL, which reduces the voltage reaching the bridge terminals. As illustrated in Figure 7.04, Kelvin connections eliminate this drop with a pair of leads added at the excitation terminals to measure and regulate the bridge voltage. For example, when ie = 50 mA, RL = 5 W, and the combined voltage drop in the two leads is 500 mV, no voltage drops in the sense wires. A commercial unit uses a Kelvin connection to measure and regulate the voltage at the bridge. It supplies the voltage to the strain gage with one pair of leads and measures it with another pair as shown in Figure 7.05. The six wires are used in pairs for Sense, Excite, and Measure. The Sense lead is a feedback loop to ensure that the Excite voltage is constantly held within specifications. Strain Gage Signal Conditioning Conductors carrying such low level signals are susceptible to noise interference and should be shielded. Low-pass filters, differential voltage measurements, and signal averaging are also effective techniques for suppressing noise interference. Furthermore, instrumentation amplifiers usually condition the extremely low strain gage signals before feeding them to ADCs. For example, a 10-V full-scale input provides 156 µV of resolution for a 16-bit ADC. The instrumentation amplifier gain should be adjusted to provide the full-scale output of the strain gage or load cell over the entire range of the ADC. Force and pressure transducers typically generate an offset output signal when no external force is applied. Instrumentation amplifiers usually contain a control to adjust this offset to zero and let the load cell cover the full range of the ADC. Most instruments also provide adjustable excitation, and gain. Common Mode Rejection Ratio EQN: 7.04. Common-Mode Rejection Ratio
By comparison, a CMRR of 115 dB introduces only 9 µV of error, which corresponds to only 0.04% of full scale. Strain gage signal-conditioning modules usually provide a regulated excitation source with optional Kelvin excitation. Onboard bridge-completion resistors may be connected for quarter and half-bridge strain gages. Instrumentation amplifiers provide input and scaling gain adjustments and an offset adjustment nulls large quiescent loads. This lets input signals use the full range of the data acquisition system and the measurements cover the full resolution of the ADC. Some strain gage signal conditioners provide fixed gain, offset, and excitation settings, but fixed settings do not take advantage of the maximum dynamic range of the ADC. It decreases the actual available resolution of the measurement. For example, many generic strain gage signal conditioner modules can be set to a fixed 3-mV/V rating. At 10 V, the excitation, offset, and gain trimming are all fixed and no adjustments can be made. An excitation adjustment lets users set the excitation voltage to the maximum allowed by the manufacturer, which maximizes the bridges output. Also, the offset adjustment lets users zero the output offset produced by either a small bridge imbalance or a quiescent deformation of the mechanical member. And the gain adjustment lets users set a gain that provides a full-scale output under maximum load, which optimizes the dynamic range of the ADC. Calibration An equation for calculating the shunt calibration resistor value is: EQN. 7.05. Shunt Calibration Resistor for Transducers
Many products include calibration software with a Windows-based program that provides several calibration methods, online instruction, and a diagnostic screen for testing the calibrated system. Transducers and Load Cells Strain Diaphragm Pressure Gages When one side of the diaphragm (called the reference pressure side) is open to the ambient atmosphere, the gage compares the inlet pressure to the ambient pressure, which is about 14.7 psi at sea level. When the gage measures ambient pressure, the reference chamber must be sealed with either a vacuum reference (near zero psi) or the sea-level reference. Temperature variations can affect the accuracy of the strain diaphragm pressure gages. A pressure gage with a sealed non-zero reference pressure exhibits temperature variations consistent with the ideal gas law. For example, a 5°C change in ambient temperature near normal room temp (25°C) produces an error of 1.7% in the pressure measurement. Temperature variations can also affect the performance of the strain gages themselves. Transducers must contain temperature compensation circuits to maintain accurate pressure measurements in environments with widely varying temperatures. All strained-diaphragm pressure gages require a regulated excitation source. Some gages contain internal regulators, so users can connect an unregulated voltage from a power supply. Some strained-diaphragm pressure gages also employ internal signal conditioning, which amplifies the mV signal output of the Wheatstone bridge to a full-scale voltage from 5 to 10 V. Gages of this type have low-impedance outputs. In contrast, other pressure gages have no internal signal conditioning so their output impedance equals the Wheatstone bridge resistance (several kW for semiconductor types), and their full-scale output is in mV. ...to read the entire 144-page book, order your copy today!
|
![]() |
|
[ IOtech PC-Based Data Acquisition (DAQ) and Instrumentation - PDF Format IOtech PC-based products (USB, PCI, Ethernet, GPIB) and signal conditioning options capture waveforms and measure most physical parameters including temperature, vibration, strain, velocity, acceleration, position, as well as common voltage, current, power, and data logging. IOtech's solutions are used in environments such as |